[. . . ] Setting the Calibration Mode Calibration Modes The CL-200 CHROMA METER offers two calibration modes: NORM and MULTI. [. . . ] The mode is appropriate for measuring a tricoloraddition light source whose spectral distribution is approximately the same as that of the reference source. Setting CFs for Each Mode The CL-200 CHROMA METER can store CFs (correction factors) for both calibration modes, and can apply these to your measurement results. You can use this feature to correct for disparities among different CL-200s or to adjust calibration arbitrarily as needed. After setting the CFs, you can use the [CF] key to switch use of CF correction ON or OFF. Setting the CFs for NORM calibration mode You set the CFs by taking a measurement of a known light source, then keying in the corresponding known values. For details, see page 26. Setting the CFs for MULTI calibration You set the CFs by measuring the three color components (R, G, and B) of the light source (a CRT or LCD projector) and then keying in the known reference values. [. . . ] For details, see page 50. Specifications and appearance are subject to change without notice. 51 9222-1877-11 ©2003-2004 KONICA MINOLTA SENSING, INC. AEMEPX(11) Printed in Japan [. . . ]